From: The functional limits of the aneurysmal aortic root. A unique pressure testing apparatus
| Porcine specimen aortic root | Maximal pressure (mmHg) | Macroscopic characteristics | Porcine specimen ascending aorta only | Maximal pressure (mmHg) | Macroscopic characteristics |
|---|---|---|---|---|---|
| 1 | 180 |
• Tissue dissection at site of pressure transducer • Circumferential spread of dissection | 1 | 300+ |
• No loss of tissue integrity • Apparatus failure |
| 2 | 200 |
• Tissue dissection at site of pressure transducer • Luminal tear at coronary ostia • Circumferential spread of dissection | 2 | 300+ |
• No loss of tissue integrity • Apparatus failure |
| 3 | 220 |
• Tissue dissection at site of pressure transducer • Luminal tear at coronary ostia • Circumferential spread of dissection | 3 | 300+ |
• No loss of tissue integrity • Apparatus failure |
| 4 | 200 |
• Tissue dissection at site of pressure transducer • Circumferential spread of dissection | 4 | 300+ |
• No loss of tissue integrity • Apparatus failure |
| 5 | 240 |
• Tissue dissection at site of pressure transducer • Circumferential spread of dissection | 5 | 300+ |
• No loss of tissue integrity • Apparatus failure |